Peristaltic pump – high flow rates
The peristaltic pump with WT300F drive and KZ25 pump head is designed for continuous liquid transfer and repeated dispensing. Versions with one or two channels are available. The liquid only contacts the inner tubing wall. In the two-channel version, both heads share a common drive; actual delivery depends on the tubing, liquid and setup.
Features and functions
- Colour touchscreen and control keys
- Four operating modes: continuous flow, volume dispensing, timed dispensing and copy dispensing
- Storage for up to five parameter groups in dispensing modes
- Flow calibration using measured test volumes
- Adjustable suck-back function to reduce dripping
- RS485/Modbus RTU for integration into an external control system; optional foot pedal
Technical specifications
| Property | Specification |
|---|---|
| Drive | WT300F |
| Pump head | KZ25; two heads in the two-channel version |
| Speed | 0.1–350rpm |
| Flow range with KZ25 | 0.31–3'850mL/min per channel, depending on tubing |
| Power supply | 230V AC, 50/60Hz |
| Power consumption | <150W |
| Protection rating | IP31 |
| Drive L × W × H | 286 × 180 × 230mm |
| Drive weight | 5.8kg |
| Communication | RS485 / Modbus RTU |
| Additional interface | RS232 according to the manual pin assignment |
Versions
| Product code | Version | Pump heads |
|---|---|---|
| 1702_1 | 1 channel | 1×KZ25 |
| 1702_2 | 2 channels | 2×KZ25 |
Flow performance and calibration
The stated flow range is per channel for the KZ25 configuration and depends on the tubing size used. It is not the sum of both channels. Manufacturer values refer to reference conditions with water at room temperature. Check and calibrate actual delivery on the installed system for your application. Adjustable dispensing volumes do not guarantee consistent accuracy for every tube and liquid.
Compatible tubing and accessories
- KZ25 pump head
- Silicone tubing #15
- Silicone tubing #24
- Silicone tubing #35
- Silicone tubing #36
- Foot pedal available separately
Compatible silicone tubing: #15, #24, #35 and #36. The offered rolls contain 20m; select the required tubing size in the linked product.
Notes on use
Check chemical compatibility between the liquid and tubing before use. Disconnect the pump from the mains before installation, tubing changes or opening the head cover. Operate the head only when fully closed. The standard version is not intended for potentially explosive atmospheres. RS485 is not an Ethernet connection; Ethernet integration requires a suitable external gateway.
Downloads
Operating instructions: Deutsch · English · Français
Modbus communication protocol (English)
Technical specifications are subject to change without prior notice.
Shipping
The shipping costs are included from 79CHF. This will be shown in advance at the time of purchase.
delivery terms
We only deliver to Switzerland and Liechtenstein. We only deliver by mail. Unfortunately, a self collection of the products is not possible.
delivery time
We keep most of our items in stock. These will be sent to your address by Swiss Post on workdays 48 hours after receipt of the order. Non-stock items are generally marked as such and will be delivered within 2 - 8 weeks. If you have any questions, our customer service is available on weekdays from 9 a.m. to 4 p.m. on +41 77 533 28 21 or at info@labmaterial.ch available.